Pump has been inactive without input for more than 10 minutes.
The Rotor Error screen appears during RUNNING or PRIMING, when the pump
has detected an unusual operating condition for the rotor. The Rotor Error
generally results from a problem with the Pump Set tubing around the rotor.
The Feed Error screen appears when the enteral formula is no longer being
delivered because the bag is empty or because of a clog between the pump
and the bag.
The Flush Error screen appears when the ushing solution is no longer being
delivered because the ush bag is empty or because of a clog between the
pump and ush bag.
The Flow Error screen appears when either the feeding or the ushing
solution is no longer being delivered because of a clog between the pump
and the patient.
The Pump Set Dislodged screen will appear if the black ring retainer (MISTIC)
is not properly loaded in the MISTIC pocket in the pump set loading area.
The Battery Low screen appears and the alarm beeps continuously when the
battery needs to be recharged. There is approximately 15 minutes of battery
life remaining when this screen appears.
The Feeding Complete information screen appears after completion of the
programmed feeding.
The pump set usage warning indicator will blink on the RUNNING screen if a
pump set has been used for 24 or more hours (hours actually running).
The pump status LED array on the upper right of the pump gives a quick
visual indication of the pump status, especially in darkened rooms.
Hold Error
Rotor Error
Feed Error
Flush Error
Flow error
Pump Set Dislodged Error
Battery Low
Feeding Complete
Pump Set Use > 24 Hours
LED Array Warning
SYMPTOM PROBABLE CAUSE CORRECTIONS
Press CONTINUE to return to the previous screen
Check that the pump set is not damaged and re-load the tubing
around the rotor as described in subsection “Loading Pump Sets”.
Press CONTINUE to restart the pump.
Check the feed line to nd the occlusion causing the blockage. If
the occlusion cannot be cleared, load a new pump set, prime it, and
press CONTINUE to restart the feeding.
Check the ush line to nd the occlusion causing the blockage. If
the occlusion cannot be cleared, load a new pump set, prime it, and
press CONTINUE to restart the feeding.
Check the line to nd the occlusion causing the blockage. If the error
cannot be resolved, check the valve pocket in the pump set loading
area for dampness or dirt, which may be a cause for a false error.
Clean and dry the valve pocket. If the error still cannot be resolved,
load a new pump set, prime it, and press CONTINUE to restart the
feeding.
Check and correct the positioning of the MISTIC retainer if possible.
If the error cannot be resolved, load a new pump set, prime it, and
press CONTINUE to restart the feeding
Plug the power cord in to begin recharging. The pump will
automatically return to the screen that was active prior to the error.
The battery will charge continuously whenever the pump is plugged
into a wall outlet. 6 hours of charging is required to fully recharge
the battery pack.
Press Power Down to turn o the pump. Press CONTINUE to return to
the SET LOADED opening menu.
It is recommended to replace pump sets after this length of usage.
This icon is only an informational message and does not require
action.
A green light status indicates normal pump operation and does not
require action.
A yellow light status indicates an informational or warning situation
for a low battery condition, the pump in HOLDING mode, or one of
the following errors: HOLD ERROR, FEED ERROR, FLOW ERROR, and
FLUSH ERROR.
A red light status indicates a critical pump error. The screen will
show one of the following error messages: SYSTEM ERROR, PUMP
SET DISLODGED, or ROTOR ERROR.