ENG7
Emissions Test Compliance Electromagnetic Environment - Guidance
RF emissions
(CISPR 11) Group 1
RF emissions
(CISPR 11) Class B
Harmonic emissions
(IEC 61000-3-2) Class A
Voltage uctuations/ icker emissions
(IEC 61000-3-3) Complies
Radiated Disturbance Immunity
(EN60601-1-2 / IEC 61000-4-3:2002) Complies
Conducted Disturbance Immunity
(EN60601-1-2 / IEC 61000-4-6:2001) Complies
Power Frequency Magnetic Field Immunity
(EN60601-1-2 / IEC 61000-4-8:2001) Complies
Voltage dips and sags Immunity
(EN60601-1-2 / IEC 61000-4-11:2001) Complies
Electrical Fast Transient / Bursts Immunity
(EN60601-1-2 / IEC 61000-4-4:2001) Complies
Electrostatic Discharge Immunity
(EN60601-1-2 / IEC 61000-4-2:2001) Complies
Surge Immunity
(EN60601-1-2 / IEC 61000-4-5:2001) Complies
The KANGAROO ePump enteral feeding pump
uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby
electronic equipment.
The KANGAROO ePump enteral feeding pump is
suitable for use in all establishments, including
domestic establishments and those directly
connected to the public low-voltage power
supply network that supplies buildings used for
domestic purposes.
Electromagnetic Conformity Declaration
The KANGAROO ePump enteral feeding pump has been built and tested according to UL60601-1, E189131 CSA-C22.2 No.601.1-M90, and EN60601-1-2 Standards.
The KANGAROO ePump enteral feeding pump is intended for use in the electromagnetic environment specied below. The user of the pump should assure that it is used in such an
environment.