ENG8
Section IV — Cleaning Procedures
Cleaning should be performed as needed. It may also be desirable to dene cleaning intervals based on knowledge of the environment in which the pump is used. Only personnel
trained in the cleaning of medical devices should perform cleaning.
CAUTION: Do not immerse pump or power cord in water or other cleaning solution; clean using a damp (not wet) cloth or sponge. Failure to follow the cleaning procedures
described herein could result in hazards to users. As with any AC powered electrical device, care must be taken to prevent liquid from entering the pump to avoid electrical shock
hazard, re hazard, or damage to electrical components.
If any of the following events occur, DO NOT USE the pump until it has been properly cleaned and serviced by personnel trained in servicing KANGAROO ePump enteral feeding
pumps:
wetting of the pump’s power cord or leakage into the pump interior during cleaning.
spillage of large amounts of formula onto the pump exterior, or any spillage onto the power cord.
General Cleaning Directions
Cleaning of KANGAROO ePump enteral feeding pumps must be performed as follows:
CAUTION: Disconnect pump from AC power source before cleaning. After cleaning, do not connect to AC power source until pump and power cord are thoroughly dry.
A mild detergent should be used for general cleaning. If necessary, the pump may be cleaned with a 10:1 water and hypochlorite mixture, however, repeated cleaning with this
solution can damage the plastic pump case. Isopropyl alcohol, applied with a damp (not wet) cotton swab, may be used for cleaning dicult-to-reach areas, however, it should
be used sparingly because repeated cleaning may damage the case.
DO NOT USE strong cleaners such as Spray Nine™*, Phisohex™*, Hibiclens™*, or Vesta-Syde™* because damage to the pump case housing can result.
Directions for Cleaning Pump Housing
Refer to General Cleaning Directions before starting.
Clean outside surface with a damp (not wet) cloth or sponge and keep pump in upright position as much as possible.
Avoid excess moisture near pole clamp area.
Do not allow cleaning solution to enter the vertical vents in the housing, underneath the handle.
Directions for Cleaning Pump Power Cord
Refer to General Cleaning Directions before starting.
Unless soiling is observed, the power cord should not be cleaned.
If cleaning of the power cord is necessary, unplug from outlet and wipe the exterior surfaces of the wall plug with a cloth dampened with isopropyl alcohol.
CAUTION: Avoid exposing power cord to excess moisture, as this can lead to an electrical shock or re hazard.
Directions for Cleaning Rotor Assembly
Refer to General Cleaning Directions before starting.
Open the blue door enclosing the pump set loading area.
Loosen rotor set screw with 5/64” (2 mm) Allen wrench and gently pull rotor o shaft. After removing rotor, avoid getting any moisture in the rotor shaft opening.
Clean rollers thoroughly with warm soapy water, or isopropyl alcohol if necessary.
Be sure all parts of rotor are completely dry before putting it back onto shaft.
To replace rotor, align set screw on hub of rotor with the attened section of the output shaft.
Push rotor into place and tighten set screw (do not over tighten).
Preventative Maintenance
This pump must be periodically re-certied to assure proper functioning and safety. The recommended service interval is at least once per year. Re-certication may be done at the
user’s Biomedical Engineering Department, an outside service, or by Tyco Healthcare Group LP Factory Service. To arrange for Tyco Healthcare Group LP Factory Service in the US,
call 1-800-448-0190. In Canada, call 1-877-664-TYCO (8926). Outside of the US and Canada, please contact your local Customer Service for more information.
If a pump malfunctions, please contact your Tyco Healthcare Group LP Representative or call Customer Service for instruction.