© 2021 TytoCare All rights reserved. 71 of 75
TytoHome General Operation User Guide
Appendix D. Electromagnetic Compatibility (EMC) Information
IMMUNITY test
IEC 60601 test
level
Compliance
level
Electromagnetic environment - guidance
Electrostatic discharge
(ESD) IEC 61000-4-2
8 kV contact 2,
4, 8, 15kV air
8 kV contact
2, 4, 8, 15kV
air
Floors should be wood, concrete or
ceramic tile. If floors are covered with syn-
thetic material, the relative humidity
should be at least 30 %.
Power frequency
(50/60 Hz) magnetic
field IEC 61000-4-8
30 (A/m) 30 (A/m) Power frequency magnetic fields should
be at levels characteristic of a typical loc-
ation in a typical commercial or hospital
environment.
NOTE UT is the AC mains voltage prior to application
of the test level.
Table 2. Declaration - Electromagnetic Immunity