Chapter 1: Introduction
1-4 Veterinary Vital Signs Monitor Operation Manual
to the EMC information provided in this manual.
WARNING! There is no defibrillator synchronization output on the monitor.
Make no connections between the monitor and a defibrillator.
WARNING! This monitor will not operate effectively on patients who are
experiencing convulsions or tremors.
WARNING! This monitor is not for home use.
WARNING! The monitor should not be used adjacent to or stacked with
other equipment. If adjacent or stacked use is necessary, the
monitor should be observed to verify normal operation in the
configuration in which it will be used.
WARNING! This monitor is not for apnea detection. The monitor has not
been tested or validated for use in apnea detection.
WARNING! Verify proper operating mode before attaching patient. See
Select the Patient Type in Chapter of Setting Up the Monitor.
WARNING! Default alarm limits are provided for convenience. Verify that
alarm limits are appropriate for given patient and condition,
and adjust according to institutional policy.
WARNING! Ensure the monitor's AC rating is correct for the AC voltage at
your installation site before using the monitor. The monitor's
AC rating is shown on the rear panel rating plate. If the rating is
not correct, do not use the monitor.
WARNING! Disconnect the AC power supply from the outlet before
disconnecting it from the monitor. Leaving the AC power
supply connected to an AC power outlet without being
connected to the monitor may result in a safety hazard.
WARNING! Do not allow any moisture to touch the AC power supply
connectors or a safety hazard may result. Ensure that hands
are thoroughly dry before handling the AC power supply.
WARNING! Do not place the monitor in the patient's bed. Do not place the
monitor on the floor.