Chapter 9: Non-invasive Blood Pressure
9-10 Veterinary Vital Signs Monitor Operation Manual
If the measurement is successful, you will see the systolic, diastolic, and mean
arterial pressure values displayed in the NIBP parameter box and you will hear
the completion beep. The time of the successful reading will also be displayed
in the parameter box. When the five minutes have passed or if you press NIBP
key to cancel, the mode will automatically return to the previous setting
(MANUAL or AUTO).
If the measurement is not successful, you will see dashes (- - -) displayed in the
place of the pressure values. You will see a message displayed in the NIBP
parameter box. NIBP readings will not continue until the error message is
acknowledged, but the five-minute timer will continue to run.
9.7.4 Canceling the an NIBP Measurement
Regardless the selected mode (AUTO, MANUAL, or STAT), you can stop a
measurement in progress by pressing the NIBP key on the monitor.
If the monitor is in automatic (AUTO) mode, another measurement will start
after the selected interval has elapsed.
If the monitor is in STAT mode and you press the NIBP key to cancel the
measurement, the mode will automatically return to the previous setting
(MANUAL or AUTO).
9.8 Cleaning the NIBP Cuff
1. Place the Cuff Cap on the end of the cuff hose.
NOTE! Failure to place the cuff cap on the hose will allow water to enter
the bladder. This will cause the cuff to malfunction and will
damage the monitor.
2. Pre-clean the cuff by removing contaminate material, hair, and debris.
3. Wash the cuff in a regular load of laundry, or hand-wash.
4. Allow the cuff to air-dry.
CAUTION! Do not place the cuff in the dryer.