5. Start the adjustment sequence from Main menu > Functions > Adjustments.
6. MI70 notifies you that automatic power o is disabled during adjustment mode, press OK
to acknowledge.
7. Select the RH parameter when prompted.
8. The adjustment mode is now active, and you can see the measured RH reading on the
screen. To proceed with the adjustment, press Adjust.
9. Select 1-point adjustment.
10. You are now in the 1-point adjustment screen. Wait until the measurement has stabilized
fully in the reference environment, and then press Ready.
Chapter 4 – Calibration and Adjustment