EasyManua.ls Logo

Vaisala HMD62 - 5.7 Setting Filtering Factor with Insight

Vaisala HMD62
52 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5.7 Setting Filtering Factor with Insight
Figure 19 Filtering Factor Configuration View (HMD62 Example)
To set a filtering factor for the transmitter's measurement output:
1. Select > Configure Device > Filtering.
2. Enable filtering with the ON/OFF selections (slide right to enable, left to disable).
3. Enter a filtering factor below 1.000 to apply filtering to the output (range: 0.001 … 1.000).
Examples of the eect of filtering on output:
Filtering factor 1.000 = No filtering, the latest measurement is output
directly without integrating previous measurements.
Filtering factor 0.500 = The reading output integrates 50% of the
previous measurement with the latest measurement.
Filtering factor 0.100 = The reading output integrates 90% of the
previous measurement with the latest measurement.
4. Select Save when done and exit with Close.
More Information
Filtering Factor (page 12)
HMD62 TMD62 User Guide M212016EN-A
34

Table of Contents

Related product manuals