5.6 Changing Pressure Compensation Settings
with Insight (HMD62)
Figure 18 Pressure Compensation Settings
By default, the pressure value used when calculating HMD62 measurements is 1013.2 hPa.
To change the pressure compensation setting:
1. Select > Configure Device.
2. To change the volatile compensation value (resets back to power-up default at device
restart), select Compensation setpoints, enter a value, and select Save.
3. To change the default compensation value that stays in use also after device restart, select
Compensation power-up defaults, enter a value, and select Save.
4. Select Close to exit the menu after you have saved the compensations.
More Information
‣
Environmental Compensation (HMD62) (page 12)
Chapter 5 – Operating with Insight PC Software