EasyManua.ls Logo

Vaisala HMD62 - 5.3 Insight Main View

Vaisala HMD62
52 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4. Wait for Insight software to detect the transmitter.
5.3 Insight Main View
3
1 2
Figure 15 Insight Main Menu and Settings
1 Select to access Insight main menu.
Configure Device: environmental compensation settings, analog output 1 and 2
settings,
filtering factor and general settings.
Export Settings: creates a text
file export of the device settings.
Calibrate: options for calibrating and adjusting RH and T output, testing and
adjusting current (mA) output levels, and restoring factory adjustments.
Communication: contains a quick access selection for restarting the device.
Factory Default Settings: restores the transmitter back to default settings, clears
any user adjustments and restores the latest factory calibration.
About Device: general device information such as serial number and software
version.
2 Select Settings to switch between the Basic Mode and Advanced Mode user modes,
change the units of parameters (metric/non-metric), enter a factory code to access
restricted functionalities, or view information about the Insight software.
3 Monitoring provides options for monitoring and recording selected parameters, and
exporting the monitoring data as a CSV (comma-separated values) file.
5.3.1 Basic and Advanced User Modes
You can switch between the Basic Mode and Advanced Mode user modes with the selections
in the Settings menu.
HMD62 TMD62 User Guide M212016EN-A