Certain functionalities are only available in Advanced Mode. The options enabled by switching
to Advanced Mode are often intended for administrative users: set the user mode according to
the requirements of the personnel that use the device. For example, measurement viewing,
calibration and adjustment options, and output scaling are all available also in Basic Mode.
5.4 Configuring Analog Outputs with Insight
HMD62 only: Note that output parameter selections made with the
DIP switches on the transmitter component board override the Insight
configuration. When using Insight to configure the output, set all parameter
selection DIP switches on the transmitter component board to the OFF position
(left) to ensure they do not cause a conflict with the Insight settings.
CAUTION!
Figure 16 Analog Output Configuration Options in Insight (TMD62 Example)
To configure analog output with Insight:
1. (HDM62 only) Check that the output parameter selection DIP switches on the transmitter
component board are switched to the OFF position.
2. Connect to Insight and select
> Configure Device > Analog Output 1[2].
3. Select the output mode (either 4 … 20 mA or O).
Chapter 5 – Operating with Insight PC Software
29