Section 9 | Data Requests and Output Formats
9 Data Requests and Output Formats
Notes:
• All commands must be confirmed using “Carriage Return” or “Enter” on the keyboard, with the
exception of the “Stop” command (#).
• All commands are echoed back by the instrument as they are typed
9.1 Stop Command
The instrument can be stopped when it is awake by sending the ‘#’ character.
Once interrupted the instrument returns a ‘>’, and waits for a further command. If no command is received
for 5 minutes then the unit will resume operation and return to its sampling pattern.
9.2 Run Commands
Performs continuous measurement at 0.5 Hz(pressure) / 10s interval (Radar)
Burst sampling pattern of 30 seconds / 1 minute
Burst sampling pattern of 40 seconds / 6 minutes
Burst sampling pattern of 40 seconds / 10 minutes
Burst sampling pattern of 40 seconds / 15 minutes
Burst sampling pattern of 60 seconds / 30 minutes
Resume previously set sampling rate
Outputs file table (list of data files)
Clears memory (use with caution!)
Reads Current Date & Time
$ICLK;DD;MM;CC;YY;HH;MM;SS
e.g.: $ICLK;03;02;20;08;14;30;00 sets clock to 14:30:00 on 3rd February
2008
Deletes specific file e.g.: $DEL;File1
Performs simple text output data contained in specific file
Performs Zmodem extraction of data contained in specific file
Outputs amount of free memory
Reads internal battery voltage levels