During Surgery
Instructions for Use
Force 2 User’s Guide 4-7
Typical Power Settings
The power level used for various surgical procedures varies considerably
with the surgeon's technique and the size of the active electrode. A needle
electrode requires less power to sustain a spark than a large ball electrode.
One surgeon may perform a procedure by electrosurgically severing
tissue with a cutting or blended waveform while another surgeon might
perform the same procedure by using a coag waveform at a lower power
level.
A general outline of typical power settings for various surgical
procedures:
Low Power – Under 30W
Neurosurgery (both bipolar and monopolar)
Laparoscopic sterilization (both monopolar and bipolar)
Vasectomies
Dermatology
Oral surgery
Plastic surgery
Medium Power – coag 30W to 70W, cut 30W to 150W
General surgery
Laparotomies
Head and neck surgery (ENT)
Major orthopedic surgery
Major vascular surgery
Routine thoracic surgery
Polypectomy
High Power – coag over 70W, cut over 150W
Thoracotomies (heavy fulguration, 70W to 75W)
Ablative cancer surgery, mastectomies, etc. (cut 180W to 200W; coag
70W to 75W).
If the proper setting is not known from personal experience, one should
set the generator at a very low setting and cautiously increase power until
the desired effect is achieved.