Input Power Source
Force 2 User’s Guide A-3
Technical Specifications
Input Power Source
Force 2-2 PCH Generator
The line frequency may vary between 45 and 65 Hz
Force 2-8 PCH Generator
The line frequency may vary between 45 and 65 Hz
Line Regulation
Force 2-2 PCH Generator
Between 85 and 135 volts input, output power into nominal load will vary
no more than 15% or 5 watts, whichever is greater.
Force 2-8 PCH Generator
Between 170 and 270 volts input, output power into nominal load will
vary no more than 15% or 5 watts, whichever is greater.
Maximum Operating Range
85-135 Vac
Nominal Operating Range
110-120 Vac
Current: Idle - 0.7A, max Power: Idle - 60W, max
Cut - 7.0A, max Cut - 600W, max
Coag - 2.9A, max Coag - 250W, max
Bipolar - 2.6A, max Bipolar - 220W, max
Maximum Operating Range
170-270 Vac
Nominal Operating Range
220-240 Vac
Current: Idle - 0.4A, max Power: Idle - 60W, max
Cut - 2.9A, max Cut - 500W, max
Coag - 1.5A, max Coag - 250W, max
Bipolar - 1.3A, max Bipolar - 220W, max