Adaptive REM Monitoring
6-2 Force 2 User’s Guide
Adaptive REM Monitoring
After the patient return electrode is applied and the initial impedance is
between 5 ohms and 135 ohms, the “adaptive” REM system measures the
contact impedance and uses it as a reference value. If the patient-to-pad
impedance increases 40% above this reference value or above 135 ohms,
the REM alarm sounds and the monopolar output is disabled. The
adaptive REM system is designed to calibrate itself with respect to each
patient.