High Frequency Risk Parameters
A-4 Force 2 User’s Guide
High Frequency Risk Parameters
Low Frequency Leakage (50-60 Hz)
Force 2-2 PCH Generator
All patient connected terminals tied together:
Force 2-8 PCH Generator
(IEC 60601-1, Section 19 for Class 1 Equipment)
Measured with radio frequency output inactive:
Bipolar RF leakage current:
< 59 mA, per IEC 60601-2-2.
Monopolar RF leakage current:
< 150 mA, per IEC 60601-2-2.
Source current normal polarity, intact chassis ground
< 10 µA
Source current normal polarity, ground open
< 25 µA
Source current reverse polarity, ground open
< 25 µA
Sink current, 135V applied, all inputs
< 150 µA
Chassis source current, ground open
< 100 µA
Type CF
Current Path
Normal
Condition
Single Fault
Condition
Earth Leakage Current 0.5 mA 1.0 mA
Enclosure Leakage Current 0.01 mA 0.5 mA
Patient Leakage Current 0.01 mA 0.05 mA
Patient Leakage Current (mains on the applied part) 0.01 mA 0.05 mA
Patient Auxiliary Current 0.01 mA 0.05 mA