11. Troubleshooting
Green Smart User Manual 129
ENGLISH
Error Code Description
equal to 200 V upon X-ray emission. (In case of less than or equal
to 200 V, replace the power source.)
- After replacing the inverter board, check for error after X-ray
emission. If the same error occurs, replace the monk tank.
H013 Error occurs when the kV feedback is greater than or equal to the
set value by more than or equal to 20 kV during X-ray emission
(kV feedback ≥ set value + 20). The number "4" appears on the
inverter board.
- After emitting X-rays again, check whether the same error
occurs.
- Check the input power (DC 300 V) of the inverter board.
- Check that the input power (DC 300 V) is changed to less than
200 V upon X-ray emission. (In case of less than 200 V, replace
the power source.)
- After replacing the inverter board, check for error after X-ray
emission. If the same error occurs, replace the monk tank.
H014 Error occurs when the mA feedback is less than or equal to the
50 % of the set value during X-ray emission (mA feedback ≤ set
value x 0.5). The number "6" appears on the inverter board.
- After emitting X-rays again, check whether the same error
occurs.
- Check the input power (DC 300 V) of the inverter board.
- After replacing the inverter board, check for error after X-ray
emission. If the same error occurs, replace the monk tank.
H015 Error occurs when the mA feedback is greater than or equal to the
150 % of the set value during X-ray emission (mA feedback ≥ set
value x 1.5). The number "6" appears on the inverter board.
- After emitting X-rays again, check whether the same error
occurs.
- Check the input power (DC 300 V) of the inverter board.
- After replacing the inverter board, check for error after X-ray
emission. If the same error occurs, replace the monk tank.
H020 Error occurs during movement from the P-axis motor's origin
position.
- Check the connection lines for the motor and origin sensor. If any
problem, replace the motor or sensor.
H021 Error occurs during movement from the rotator-axis motor's origin
position.
- Check the connection lines for the motor and origin sensor. If any
problem, replace the motor or sensor.
H024 Error occurs during movement from the CT horizontal laser
up/down-axis motor's origin position.
- Check the connection lines for the motor and origin sensor. If any
problem, replace the motor or sensor.
H027
Error occurs during movement from the Cephalo sensor motor's