11. Troubleshooting
130 Green Smart User Manual
Error Code Description
origin position.
- Check the connection lines for motor and origin sensor. If any
problem, replace the motor or sensor.
H030 Error occurs during movement from the 4-axis collimator's left
origin position.
- Check the connection lines for the motor and origin sensor. If any
problem, replace the motor or sensor.
H031 Error occurs during movement from the 4-axis collimator's right
origin position.
- Check the connection lines for the motor and origin sensor. If any
problem, replace the motor or sensor.
H032 Error occurs during movement from the 4-axis collimator's up
origin position.
- Check the connection lines for the motor and origin sensor. If any
problem, replace the motor or sensor.
H033 Error occurs during movement from the 4-axis collimator's down
origin position.
- Check the connection lines for the motor and origin sensor. If any
problem, replace the motor or sensor.
H036 Error occurs during movement from the 1-axis collimator's left/right
origin position.
- Check the connection lines for the motor and origin sensor. If any
problem, replace the motor or sensor.
H037 Error occurs during the generator tilting motion.
- Check the connection lines for the motor and origin sensor. If any
problem, replace the motor or sensor.
H038 Error occurs during the temple support motor's motion.
- Check the connection lines for the motor and origin sensor. If any
problem, replace the motor or sensor.
H039 Error occurs during the movement from the X-axis motor's origin
position.
- Check the connection lines for the motor and origin sensor. If any
problem, replace the motor or sensor.
H040 Error occurs during the movement from the Y-axis motor's origin
position.
- Check the connection lines for the motor and origin sensor. If any
problem, replace the motor or sensor.
H060 Error occurs when the Exposure Switch has been pressed when
the equipment is turned on.
- Check the connection to the Exposure Switch and perform Reset
the equipment.
H102 Error occurs when there is no response to CAN communication.