EasyManua.ls Logo

Vatech GREEN SMART - Page 176

Vatech GREEN SMART
200 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
15. Appendices
168 Green Smart User Manual
CBCT Mode
Test Condition
Item Condition
Tested Mode CBCT (10x8.5), High Resolution
Distance from focal point [m] 1 ~ 2
Applied Tube Voltage Peak [kVp] 99
Applied Tube Current [mA] 12
Results
Direction [°]
FOV 10x8.5, 16.4 s [mR/hr]
1 m
(3.3 ft)
1.5 m
(4.9 ft)
2 m
(6.6 ft)
0
361
85
51
45
200
112
74
90
173
109
68
135
206
102
55
180
13
9
6
225
260
85
57
270
291
106
50
315
287
110
55

Table of Contents

Related product manuals