15. Appendices
Green Smart User Manual 169
ENGLISH
15.3 Electromagnetic Compatibility (EMC) Information
Guidance and manufacturer’s declaration - electromagnetic emissions
The PHT-35LHS is intended for use in the electromagnetic environment specified as
below. The customer or the user of the PHT-35LHS should assure that it is used in
such an environment.
Immunity test Compliance
Electromagnetic
environment
- Guidance
RF emissions CISPR 11 Group 1
The PHT-35LHS uses RF
energy only for its internal
function. Therefore, its RF
emissions are very low and
are not likely to cause any
interference in nearby
electronic equipment.
RF emissions CISPR 11 Class A
The PHT-35LHS is suitable
for use in all establishments
other than domestic and may
be used in domestic
establishments and those
directly connected to the
public low-voltage power
supply network that supplies
buildings used for domestic
purposes, provided the
following warning is heeded:
Warning: This
equipment/system is intended
for use by healthcare
professionals only. This
equipment/ system may
cause radio interference or
may disrupt the operation of
nearby equipment. It may be
necessary to take mitigation
measures, such as re-
orienting or relocating the
PHT-35LHS or shielding the
location.
Harmonic emissions IEC
61000-3-2
Class A
Voltage fluctuations / flicker
emissions IEC 61000-3-3
Complies
NOTE) It is essential that the actual RF shielding effectiveness and filter attenuation
of the shielded location be verified to ensure that they meet or exceed the specified
minimum values.

Table of Contents