EasyManuals Logo

Vatech GREEN SMART User Manual

Vatech GREEN SMART
200 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #99 background imageLoading...
Page #99 background image
8. Acquiring CEPH Images (Optional)
Green Smart User Manual 91
ENGLISH
6. Align the Ear Rods to the patient’s ears properly so that the head does not move during
the operation. Moreover, align the Nasal Positioner with the patient’s nation by adjusting its
height.
7. Align horizontally, so the patient's Frankfurt Line is parallel to the floor.
8. Direct the patient to swallow first before closing the mouth and to remain in his/her current
position until image acquisition is completed.
9. Click the READY button. The x-ray exposure has not started yet.
10. Now go to 8.4 X-ray Exposure to start the exposure.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Vatech GREEN SMART and is the answer not in the manual?

Vatech GREEN SMART Specifications

General IconGeneral
BrandVatech
ModelGREEN SMART
CategoryMedical Equipment
LanguageEnglish

Related product manuals