EasyManuals Logo

Vatech PAX Duo3D User Manual

Vatech PAX Duo3D
74 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #68 background imageLoading...
Page #68 background image
66
PaX-Duo3D
Chapter 8. Technical specification
8.1. PaX-Duo3D Specifications
z Weight: 185 kg
z Measurements (mm): (L) 1490*(W) 1057*(H) 2346
z X-ray Beam: Cone Beam
z Data Bit: 12bit
z Slice Thickness (mm) range: 0~10mm
z Rotating Unit Scan Angle (degree): 185 deg for 5 x5, 360 deg for other FOVs
z Number of Views: 720 frames
z Number of Sliced Images: 384 slices
z Voxel size: 0.2 mm
z Patient Position: standing
z Patient Alignment: Normal standard Arch Alignment
z FOV: 5 x5, 8.5x5, 8.5 x8.5, 12x8.5 (Cm)
z Reconstruction Time: 5 x5:54 sec, 8.5X5: 58 sec:
8.5x8.5: 60 sec, 12x8.5: 86 sec
z Exposure Time:
Standard Panoramic Adult/Child
Standard scan 13sec
High scan 9sec
Hemi-Panoramic (Left and Right) 6.5sec
Frontal Dentition 10sec
TMJ Open/Close mouth 2.8x4sec
Maxillary Sinus 10sec
Incisor clear 5sec
Canal clear 3.8sec
Maxillary Molar clear 3.8sec
z Scan Time: High 24 sec, Normal 15 sec
8.2. Panorama X-ray detector
z Technology: TDI type (Time Delay Integration)
z Active Area: 147.456x6.144mm
z Image Acquisition Area: 146x6mm
z Pixel Resolution: 5.2 lp/mm
z Gray Level: 12 bit