48
6 Set up with the display and adjustment module
FIBERTRAC 31 • Probus PA
43833-EN-170126
Caution:
When the PIN is active, adjustment via PACTware/DTM as well as
other systems is also blocked.
In delivery status, the PIN is 0000.
Call our service department if you have modied and forgotten the
PIN.
Additional settings
When a reset is carried out, all settings (with only a few exceptions)
are reset. The exceptions are: PIN, language, SIL and HART mode.
The following reset functions are available:
Basic settings: Resetting of the parameter adjustments to default
values at the time of shipment. Order-specic settings are deleted.
Default settings: Resetting of the parameter adjustment like under
"Basic settings". In addition, special parameters are reset to default
values. Order-specic settings are deleted.
Peak values of measured value: Resetting of the parameter adjust-
ments in the menu item "Setup" to the default values of the respective
instrument. Order-specic settings remain but are not taken over into
the current parameters.
Peak values of temperature: Resetting of the measured min. and
max. temperatures to the actual measured value.
The following table shows the default values of the instrument. The
values apply for the application "Summation Slave". The application
must be selected rst.
Depending on the instrument version, not all menu items may be
available or they may be dierently assigned:
Menu Menu item Default value
Setup Sensor address 126
Measurement loop name Sensor
Isotope Cs-137
Application Summation Slave
Outputs Deactivated
AI FB1 Channel Process value
AI FB1 scaling unit %
AI FB1 scaling
0 % = 0 %
100 % = 100 %
AI FB1 damping 0 s
Lock adjustment Released
Additional settings -
Reset