18 Installation
4.2 Installation kit for the wall mounting
A number of kits are available for installing a wall
mounting, e.g. on a panel inside a vehicle. The kit
size varies according to the number of modules at-
tached to the MEDUMAT Standard.
Please refer to the sheet enclosed with each installa-
tion kit for details of dimensions and installation
procedure.
4.3 Connecting up the oxygen cylinder
First wash your hands thoroughly before starting to
work on the oxygen supply. Products containing
hydrocarbons (e.g. oils, greases, alcohols, hand
creams, sticking plasters) can cause explosive reac-
tions if they come into contact with high-pressure
oxygen.
Never use wrenches or similar tools to tighten
or loosen the screw connections.
Removal of empty cylinder
1. Close the valve of the oxygen cylinder.
Switch on MEDUMAT Standard with ON/OFF
switch. This exhausts any residual oxygen and
depressurizes the ventilator. Wait until the pres-
sure gauge on the pressure reducer shows
0 bar
oxygen content before uncoupling the screw
connection by hand.
2. First switch off MEDUMAT Standard again.
3. Then loosen the screw connection to the
cylinder.