28 Using the ventilator
EN
5.8 Terminating ventilation
Important note!
Never empty the oxygen
cylinder completely. Re-
turn the cylinder for filling
while it still contains resid-
ual pressure. This prevents
entry of moist atmospheric
air that can cause
corrosion.
1. Check the oxygen supply on the gauge on the
pressure reducer gauge. If the pressure has
dropped to 50 bar or less, the cylinder should be
refilled or a reserve cylinder should be available in
order to ensure that the ventilator will be able to
function properly.
2. Close the valve of the oxygen cylinder.
3. Switch off MEDUMAT Standard. ON/OFF switch
must be kept pressed down for at least
2 seconds until the LEDs in the alarm panel light
up. This is a safety device to prevent the ventila-
tor from being switched off unintentionally.
5.9 Alarm signals
Alarm panel signals the following alarms:
Stenosis: Stenosis or a rise to maximal ventila-
tion pressure P
max
in two successive
inspiration phases
Disconnection:Disconnection between
MEDUMAT Standard and the patient
in two successive inspiration phases
< 2,7 bar O
2
: Drop in oxygen pressure to below
2.7 bar
: Battery charge inadequate
All the optical alarms are accompanied by an
acoustic alarm.
Stenosis
Disconnection
< 2,7 bar O
2