32 Using the ventilator
5.11 Alternative ventilation procedures
If MEDUMAT Standard ceases to function during a
ventilation procedure, the following alternatives can
be applied:
Ventilation bags
1. Remove the patient valve from the tube or the
mask.
2. Replace it with the ventilation bag, e.g. a WEIN-
MANN Emergency COMBIBAG WM 11000, and
perform manual ventilation.
Exhaustion of oxygen supply
In emergency situations when the oxygen supply
runs out, MEDUMAT Standard can also function
with compressed air.