Functional checks 39
EN
7. Functional checks
Prior to each use, after each dismantling and reassem-
bly and at the very least every six months, the user
must carry out functional checks on the ventilator.
N.B. Before carrying out the functional check on
MEDUMAT Standard, you must connect up the ven-
tilation tube and the patient valve.
MEDUMAT Standard must not be used if the func-
tional checks reveal defects or deviations from the
selected parameters.
First try to correct the error with the help of the in-
formation provided in section “8. Troubleshooting“
on page 48. If this is not possible, have the device re-
paired by the manufacturer WEINMANN Emergency
or by specialists explicitly authorised to do so by
same.
A complete functional check includes:
“7.2 Check for leaks in the system“ on page 41
“7.4 Checking the minute volume“ on page 43
“ Check the breath volume“ on page 43
“7.5 Check of maximal ventilation pressure“ on
page 44
“7.6 Check of alarm systems“ on page 46
We recommend that you hold reserve stocks of the
following items:
washers for the connections,
dust filters,
lip membranes for the patient valve.