Remove static-sensitive components and assemblies from their static-shielding bags
only at static-safe workstations—a properly grounded table and grounded floor mat—
and only when you are wearing a grounded wrist strap (with a resistor of at least 1
megohm in series) or other grounding device.
Use only grounded tools when inserting, adjusting, or removing static-sensitive
components and assemblies.
Remove or insert static-sensitive components and assemblies only with system
power turned off.
Insert and seal static-sensitive components and assemblies into their original static-
shielding bags before removing them from static-protected areas.
Always test your ground strap, bench mat, conductive work surface, and ground
cord before removing components and assemblies from their protective bags and
before beginning any disassembly or assembly procedures.
Service manual Safety 7