8. Remove the wall anchor screws using a #2 Phillips screwdriver.
9. Hold the bottom and top of the system and push it up to remove the system from
the wall mounting bracket.
10. Place the wall system on a table or flat work surface so that the back of the wall
system faces up. Ensure that the display is protected from pressure or getting
scratched.
11. Locate the battery, indicated by
12. Remove the battery.
Remove the light-bar board
Complete the following preliminary steps:
•
Power down the system and disconnect the AC power cord.
• Remove the system from the wall and remove the battery.
1. Remove the two torx screws from the light-bar cover.
2. Remove the light bar and cover.
110 Section B Welch Allyn Connex
®
Integrated Wall System