Configuration Manager messages
Message Possible cause Suggested action
Alarm
Unable to load configuration; using
factory defaults.
A configuration load error occurred Restore factory defaults. If the
error persists, replace the main
board.
Functional error. Call for service. A critical configuration load error occurred Replace the main board.
Information
No connection for send. The monitor is not configured to the network
In Advanced settings, touch the
Data Management > Clinical
Data tabs, then disable
Automatically send on
manual save if you only want to
store readings on the device.
If your want to send readings,
then configure the device to the
network.
Patient data management messages
Message Possible cause Suggested action
Alarm
Maximum number of patient
records saved. Oldest record
overwritten.
The maximum number of patient records has
been saved on the device
Go to the Review tab and delete
old records to prevent the alarm
from appearing when new
records are saved.
Information
No data saved. No patient data is available Take or enter vital signs before
saving.
Patient ID required to save data. The configuration requires a patient ID to save
data
In Advanced settings, touch the
Data management > Patient
tabs, then disable Require
patient ID to save readings.
Clinician ID required to save data. The configuration requires a clinician ID to save
data
In Advanced settings, touch the
Data management > Clinician
tabs, then disable Require
clinician ID to save readings.
Patient ID required to send data. The configuration requires a patient ID to send
data
Add a patient ID.
Patient list is full. Delete some
patients to add more.
The maximum number of patients was exceeded Delete a patient from the list to
add a new patient.