Note The modules fit tightly in the housing but can become loose
over time. Support the modules as you remove the module
cover.
5. Remove the modules.
Slide each module out of the housing by inserting a finger in the slot at each end of
the module and gently pull the module away from the housing until you completely
remove it.
6. Disconnect the USB cables from the rear of each module.
Reassembly notes
When aligning the module with the housing, the USB connection is on the left.
The NIBP module must always be in the front-most slot.
Remove the module well
Option 1. Complete the following preliminary steps if your goal is to replace the front
housing:
Power down the system and disconnect the AC power cord.
Remove the system from the wall and remove the battery.
Remove the handle module assembly.
Open the chassis.
Remove the light bar.
Remove the power panel.
Remove the power wire harness and the main wire harness.
Remove the modules in the bottom housing.
Option 2. Complete the following preliminary steps if your goal is to replace the power
button: