Reassembly notes
For ease of assembly, connect the IR wire assemblies to the optical sensors before
attaching the handle cradle to the handle assembly housing.
Attach the connector with the black and white wires to the optical detector
(indicated by the letter "B" molded into the cradle plastic). Attach the connector with
the red and white wires to the optical emitter. The wires should face inward once
the handle cradle is assembled.
Snap the handle cradle assembly into the front housing. Make sure that the cradle
assembly tab and the beveled edge are at the top of the housing.
Open the chassis
Complete the following preliminary steps:
Power down the system and disconnect the AC power cord.
Remove the system from the wall and remove the battery.
Remove the handle module assembly.
If the Braun option is accessible from the side, remove it as described in "Replace
modules from the main housing."
1. Complete one of the two substeps below based on your system details.
Service manual Section A 75