Activate licenses
Licensed features are enabled using the licensing application in the Welch Allyn Service
Tool. When you purchase a licensed upgrade for the Connex device, you receive an
authorization code from Welch Allyn Customer Service. This code enables you to
activate the new features through Service Tool's Install License feature.
1. Connect the Connex IWS to a PC running the Service Tool with a type A to type B
mini USB cable.
2. Launch the Welch Allyn Service Tool.
3. If the startup screen with the “Add new features” and “Service” buttons appears,
click Add new features.
4. Either log in as ADMIN with no password or use any account previously created.
5. Click on the Administration menu and select Install license.
6. Select Medical device license for the type of license you want to activate.
7. Enter your authorization code.
8. Select the device from the Device list.
Note The device must either be connected or on the maintained
device list to appear.
9. Click Activate.
Note If the license you are activating is for a connected device, the
license is installed immediately. If the license is for an offline
(maintained) device, the license will be installed the next
time the device is connected to the Service Tool. A new
license activation action will be added to your work list.
10. After installing the license, power down and restart the Connex device to load the
license and activate the new licensed features.
Note If a forced shutdown occurred immediately after installing
the license, you might need to reinstall the license.
11. If you observe any issues or errors in the activation process, confirm that you
entered the correct authorization code, check your Internet connection, and repeat
the installation process. Confirm with your systems administrator that ports 5093
and 5094 on your firewall are open for TCP/IP communications. If you cannot resolve
the issue, contact Welch Allyn Technical Support.
Note Masimo SpHb requires an additional license that is installed
as a software upgrade using the Service Tool. Additional
instructions for installing the Masimo licenses are included
with the Masimo parameter upgrade kits.
Host firmware requirements
This section lists the host firmware requirements for each parameter. If your device does
not meet the minimum requirements, you can upgrade the host firmware with the
Welch Allyn Service Tool.
SpHb requirement: host firmware version 1.70.03 or later
Braun ThermoScan PRO 6000 requirement: host firmware version 2.30.00 or later
170 Options, upgrades, and licenses Welch Allyn Connex
®