Settings Default value
Enable pulse timer Enabled
Enable Celsius only selection Disabled
Pulse rate
Settings Default value
Alarms
Alarm limits on/off control On
Upper limit Adult: 120 bpm
Pediatric: 150 bpm
Neonate: 200 bpm
Lower limit Adult: 50 bpm
Pediatric: 50 bpm
Neonate: 100 bpm
Setup
Tone volume Off
Advanced
Display source Enabled
RR
Settings
Default value
Alarms
Alarm limits on/off control On
Upper limit (etCO2) Adult: 30 breaths per minute
Pediatric: 40 breaths per minute
Neonate: 65 breaths per minute
Upper limit (ECG) Adult: 30 breaths per minute
Pediatric: 40 breaths per minute
Lower limit (etCO2) Adult: 5 breaths per minute
Pediatric: 10 breaths per minute
Neonate: 25 breaths per minute
Lower limit (ECG) Adult: 5 breaths per minute