2. Launch the Welch Allyn Service Tool.
3. In the startup screen, click Service.
4. Either log in as ADMIN with no password or use any account you previously created.
5. Select the model name of your device (CVSM or CIWS) from the Device list tab.
The Device Information tab opens.
6. Locate the host controller in the Device Name column and follow the row across to
the Available firmware column to identify the version currently running on your
device.
Install options
Install internal options
All internal option installations entail opening the device case and performing some
disassembly. Whenever you repair, replace, or install internal components, Welch Allyn
requires that the device undergo a full functional test and electrical safety test before
placing the device back in service.
Before installing an internal option, read the information about removing the option in
“Disassembly and repair.” After familiarizing yourself with the process, follow the
instructions in the disassembly section to remove currently installed components or
empty housings as needed, and then proceed to install new options.
Install external options
External options include ECG, Braun thermometers, barcode readers, and scales.
Perform any required software upgrades and install any required licenses before
connecting the option to the Connex device. Follow any instructions included with the
option to set up or mount the option before connecting it to the Connex device.
1.
Power down the Connex IWS and disconnect the AC power cord.
2. Loosen the captive retention screws on the communications cover housing and
remove the cover from the wall system.
3. Loosen the two screws on the cable retention clamp and remove it.
4. Plug the USB connector of the module external option (e.g., barcode reader) into any
available USB port. Check that the connector is fully inserted.
5. Thread the cable through the cable guide.
6. Replace the cable retention clamp and tighten the two screws.
7. Re-attach the communications cover and tighten the two retention screws.
8. Connect the AC power cord to the device.
9. Power up the wall system.
10. After the device boots up, follow the instructions in the device's and/or option's
directions for use to configure the option as necessary.
Upgrade software
If the current software version does not meet the minimum required version, update the
Connex IWS firmware to the latest version that supports the option. To update the
168
Options, upgrades, and licenses Welch Allyn Connex
®
Integrated Wall System