6. If SureTemp is installed, remove the module and the probe cover storage together,
sliding them out the open end, and disconnect the USB cable.
7. Slide any blank face plates or modules to be removed toward the open end.
Note When removing any connected modules, disconnect the
USB cables before you remove the module.
Note If you are replacing a Braun 4000 thermometer with a Braun
6000 thermometer, you must also replace the dock.
Remove the Braun dock from the housing
1. Hold the Braun housing with your right hand and the dock with your left hand as
shown. With your right-hand fingers inside the housing, gently press the housing
toward you and away from the dock. At the same time, gently push the dock away
from you until the dock unsnaps from the housing.
2. When the dock is completely loose, remove it from the housing.
Reassembly notes
• When aligning the module with the housing, the USB connection is on the bottom.
• The temperature module must always take up the last two spots closest to the end.
• The blank cover's beveled edge must be on the right when facing the device for
proper alignment.
Service manual Section A 67