Safety
Operating Manual MXPC 303 Z/603 T 15
Equipment
CAUTION
Cold surfaces: do not touch, risk of injury!
If an extremely cold coolant is used, surfaces such as those on the emission
condenser, can reach dangerously low temperatures and lead to frostbite.
The operator must ensure safe operation and take precautions to protect the
user, where necessary (see chapter 2.3 Safety precautions).
2.4.5 Vacuum
CAUTION
Risk of injury due to explosion!
The vacuum equipment can explode due to a rise in pressure after a sudden
intake of air. The splinters can seriously injure the user.
Check the pressure of the vacuum pump before disconnecting the pneumatic
connections.
NOTE The vacuum equipment can be under vacuum even when it is switched off.
GLASSWARE
Glassware is particularly high risk.
Make sure that glassware is shatter-proof and unbreakable!
Glassware must be suitable for the intended application.