Description
22 Operating Manual MXPC 303 Z/603 T
4.3 Attachment parts
Beware of the safety/warning instructions according to chapter 2.4.5 Vacuum.
To ensure safe operation, the attachment parts must always be used (see
chapter 2.3.2 Safe operation).
Side separator inlet
The separator connected upstream on the suction side protects the diaphragm pump.
Condensate and liquid media are collected in the catchpot.
Exhaust vapour condenser
The exhaust vapour condenser filters out the residual vapours and catches them in the
separator at the exhaust.
The device's suction connection must not be closed during operation (see Fig.
4-1/10 and Fig. 4-2/10).
Safety valve
The gas inlet on the emission condenser contains a safety valve. It opens in case of
overpressure, e.g., if it is clogged by deposits. Check the valve seals at regular intervals.
4.4 Accessories
NOTE The scope of delivery is defined by the supply contract.
4.4.1 Accompanying material
Optional suction port, 2nd application
Connector for hose
Compression fitting
See chapter 3.3 General Connections
Vacuum controller VCpro 601 table-top device
Measuring/control device
Working range 1100 – 1 mbar
Regulation via vacuum control/vent valve
Operating modes: manual, automatic and
programme
Vacuum pressure gauge with swivel adapter
Measuring range 1000 – 1 mbar
360° display
G1/4" threaded connection
Vacuum pressure gauge (digital) with swivel
adapter
Measuring range 1000 – 1 mbar
360° display
G1/4" threaded connection