Installation LF 170
80
ba54104d09 11/2010
3.3.2 Line power connection
WARNING
If the power supply is connected incorrectly, it may represent a
danger to life from electric shock. Pay attention to the following
points during installation:
z The monitor must only be connected by an electrician.
z The monitor must only be connected in a voltage-free
condition.
z The power supply must fulfill the specifications given on the
nameplate and in chapter 8 T
ECHNICAL DATA.
z A switch or circuit breaker must be provided in the building
installation as an interrupt facility for the monitor. The interrupt
facility must be
installed close to the monitor and easy to reach by the user
and
marked as an interrupt facility for the monitor.
z The monitor must only be opened after installation if the mains
voltage has been switched off beforehand.
The correct voltage is given on the nameplate (see section
1.3 N
AMEPLATE). In all cases check that the correct mains voltage is
present prior to installing the instrument.