EasyManua.ls Logo

wtw LF 170 - Setting Table of Parameterization

wtw LF 170
68 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation LF 170
96
ba54104e09 11/2010
4.5.2 Setting table of parameterization
Overview of the settings
(main level)
x Menu appears for the corresponding instrument variant
1)
For relays that are configured as limit indicators
2)
Menu only appears if the conductivity ( ) measuring mode is selected
General operating
instructions
The starting point for the settings is the start display of the
parameterization level (see section 4.5.1):
Operation:
z Navigate in the main level to the next or previous setting with
<UP><DOWN>. The entry level with further settings is reached by
subsequently pressing <ENTER>.
z In the entry level select the desired value from the selection in each
case with <UP><DOWN> and confirm this with <ENTER>. This puts
the setting into effect and you go to the next setting or return to the
main level.
z Certain settings or choices depend on the instrument variant or
specific configuration settings. Settings that are not available will be
Display Description
Instrument version
Stan-
dard
RT RT RS
Pr1 Current output 1 x x x
Pr2 Current output 2 x x
PL Relay limits
1)
xx
PIF RS 485 interface x
PCd Password protection x x x
Pt Temperature balancing x x x
Ptr Reference temperature
2)
xxx
PtF Temperature compensation
2)
xxx
PCS Timer function x x
Switch to the first setting of the
main level with <ENTER>.
AP

Related product manuals