Operation LF 170
92
ba54104e09 11/2010
CR1
(continued)
The following settings are possible:
, Measuring range 0.000 ... 1.999 µS/cm:
(fixed range)
, Measuring range 0,00 ... 19,99 µS/cm:
(step size 1 µS/cm)
, Measuring range 0.0 ... 199.9 µS/cm:
(step size 10 µS/cm)
, Measuring range 0 ... 1999 µS/cm:
(step size 100 µS/cm)
, Measuring range 0.00 ... 19.99 mS/cm:
(step size 1 mS/cm)
(continues on the next page)
Setting Selection/Values Explanation
(values in bold typeface = values on delivery)
Upper final value: 0,000 µS/cm
Lower final value: 1.999 µS/cm
Upper final value: 2.00 ... 19.99 µS/cm
Lower final value: 0.00 ... 15.00 µS/cm
Minimum spacing: 2 µS/cm
Upper final value: 20.0 ... 199.9 µS/cm
Lower final value: 0.0 ... 150.0 µS/cm
Minimum spacing: 20 µS/cm
Upper final value: 500 ... 1999 µS/cm
Lower final value: 0 ... 1500 µS/cm
Minimum spacing: 500 µS/cm
Upper final value: 5.00 ... 19.99 mS/cm
Lower final value: 0.00 ... 14.99 mS/cm
Minimum spacing: 5 mS/cm