press repeatedly until the cursor is flashing under ZERO
confirm (Direction indicator flashing) and the previously stored ZERO offset is
displayed.
place the instrument on the designated position
press HOLD key to register the first measuring value.
The display shows -oooooo- for a few seconds. Then the measured value is
displayed and the cursor under ZERO is flashing again.
Turn the instrument by 180°
press again to register the second measuring value. The display shows -
oooooo- for a few seconds. Then the measured value is displayed and the
direction indicator on the display is flashing.
press ENTER, to accept the correction factor for this reversal measurement
Displayed mode changes to ABSOLUTE
2.2.4 LEVELTRONIC “NT” 41 with LEVELMETER C25/DC (red cover) or LEVELMETER 2000
The LEVELTRONIC „NT“ 41 can only be used in combination with a
LEVELMETER C25/DC (red cover) Art. no. WYLER: 025-005
or a LEVELMETER 2000. Art.-no. 065-004-001.
It is also possible to connect two instruments to the mentioned
Levelmeter for differential measurements.
LEVELTRONIC „NT“ 41
with LEVELMETER 2000
Inserting the battery.