ATF System with C410:V3 Controller
XCell
TM
ATF System cycle change can occur by either of three algorithms. The screen tracks
cycle change Algorithm performance. The three Algorithms are:
1. Set Point - normal process based on set up parameters.
2. Overtime - when P2 Set Point is not achieved in within 120% of calculated time.
3. Overflow - When P2 reaches P1 (+/- OSP) within the set delay time.
The Batch Info Algorithm screen will display parameters in the following tables:
Field
Description
Algorithm
Performance
Tracks cycle change mechanism
Primary Method
Shows cycle count triggered by Set point method, tracks Pressure
cycle and Exhaust cycle counts
Overtime
Shows cycle count triggered by Overtime, tracks Pressure cycle and
Exhaust cycle counts
Overflow
Shows cycle count triggered by Overflow, tracks Pressure cycle and
Exhaust cycle counts
Button
Description
Reset Count
Resets all counts, by Set Point, Overtime and Overflow to zero
Overview
Navigates to Batch Overview Screens
All Primary
All Primary Screen Buttons are displayed to navigate to those screens
G.2
ALGORITHM