SonTek, a Xylem brand
RiverSurveyor S5/M9 System Manual (February 2013)
149
Figure 132. Loop Correction File
4. The Loop Method results will be processed and displayed as shown in Figure 133. Click
Close after the results are reviewed.
Figure 133. Loop Method Results
5. The moving bed speed is now calculated. Prior to applying the correction, you must first
open the Processing Toolbox by clicking the Tools icon (Ctrl+T) on the tool bar located at
the top of the screen. It is also helpful to open the Discharge Summary (Ctrl+S) by clicking
the summary icon located on the toolbar. The screen should appear as shown in Figure
134.