There is only English format for parameter print, the explanation is as follows
Content of print Explanation
C602 Ver1.00 Instrument type and software version No.
Max=: 020.000kg Max. capacity 20kg
e= : 01 Scale interval (effective bit) 1
Dp : 03 Decimals 3
0_AD: 00262121 AD code of calibrated zero point
0Point: 000.000kg Current zero point (corresponding to calibrated zero
point)
R : 00067106 Standard rate
Line: 000.000%FS Non-linear modification value
COMM:111110 Communication setting, corresponding toSET 1
parameter 1
0_SET 141 Zero-settingrelative toSET 1parameter 2
Flt : 1222 Filtration setting, corresponding to SET 1
parameter 3
Addr: 01 Communication addresscorresponding toSET 1
parameter 4
Buad: 44 Baud ratecorresponding to SET 1parameter 5
Mode: 22 Communication mode, corresponding toSET 1
parameter 6
Prnt: 1 Printer type, corresponding toSET 1parameter 7
PL : 00 Print language, corresponding toSET 1parameter
8
AutoP 12 Automatic print, corresponding to SET 1
parameter 9
Aout_W 020.000kg Full measurement range weight of analog output,
corresponding toSET 1parameter 10
Aout_0 00000 Zero DA code of analog output, corresponding to
SET 1parameter 11
Aout_F 65070 Zero DA code of analog output, corresponding to
SET 1parameter 12
Type: 212 Equipment type, corresponding to SET 1
parameter 13
Disp: 04 Display content, corresponding to SET 1
parameter 17
Light: 5 Display brightness, corresponding to SET 1
parameter 16