2-inch Pipe Mounting
2-inch pipe
Mounting fixture
Washer
Clamp screw
Figure 3.4.4  2-inch Pipe Mounting
1.  Pass the four clamp screws through the mounting 
xture.
2.  position it on the 2-inch pipe, and then fasten the 
AXG1A in place.
  Panel Mounting
Panel cutout
172 (6.8)
R3MAX
Mounting fixture
Washer
Screw
Clamp screw
203 (8.0)
Figure 3.4.5  Panel Mounting
1.  Fit the AXG1A into the panel.
2.  Attach the mounting xture to the AXG1A using the 
screw and the washer, and secure the product with 
the two clamp screws.
3.4.4  Mounting of AXFA11 Transmitter
AXFA11 Remote Transmitter can be mounted using 
surface mounting, 2-inch pipe mounting, or panel 
mounting, by the same process of AXG1A. Please refer 
3.4.3. for the process. 
IMPORTANT
Mounting xture on equipment intended to be mounted 
on a wall or ceiling shall withstand a force of four times 
the weight of the equipment (AXFA11: 3.4 kg (7.5 lb)).
3.5  Changing Direction of  
Cable Entry
IMPORTANT
The following types can not be changed direction of 
cable entry after delivery.
•  Submersible Type (Sufx code: -W)
•  For District Heating and Cooling or Condensation-
proof (Optional code: DHC)
(1) The following tool is required.
•  Hexagonal wrench (nominal size 5)
(2) Turn off the power to the owmeter.
(3) Using the wrench, unscrew the four hexagonal bolts 
at the neck of transmitter or terminal box.
AXW Remote SensorAXW Integral Flowmeter
CAUTION
Do not loosen the screw at the opposite side of cable 
entry (see the gure below).
F0331.ai