<5.  Basic Operating Procedures>
48
IM 01E24A01-01EN
5.  Basic Operating Procedures
5.1  Operation by Display unit
The parameter settings from display unit can be carried 
out using the three IR (infra-red) switches - namely, the 
[SET] [SHIFT] and [▼] switches. The IR switches enable 
the user to set parameters from the outside of the glass of 
the display cover.
This section provides descriptions of basic parameter 
conguration and operation procesures of IR switches. 
This product can be also operated using the dedicated 
handheld terminal or the FieldMate (Versatile Device 
Management Wizard). For operation in details, read the 
user’s manual of the applicable communication type 
(for AXW/AXW4A) or the hardware/software edition (for 
AXFA11) as listed in Table 1.1.
WARNING
 
Be sure to enable the write protect function to prevent 
the overwriting of parameters after nishing parameter 
setting. 
In rare cases, the IR switches may respond 
unexpectedly to water drops or extraneous substances 
sticking on the surface of display panel, due to the 
operating principal. The possibility of malfunction arises 
after rain or cleaning operation near the place where the 
owmeter is installed. Turning on and off the ashlight 
etc. towards the IR switch may also be a cause of 
malfunction.
Read Section 6.3 for the hardware write protect 
function, and the user's manual of applicable 
communication type as listed in Table 1.1 for the 
software write protect function.
IMPORTANT
Operate the display unit under the condition where 
direct sunlight, etc... do not shine to the IR switches 
directly when the parameter setting operation is carried 
out.
NOTE
•  Always keep the cover closed and operate the 
setting switches from the outside of the glass 
window.
•  If dirt, dust or other substances surfaces on the 
glass of display cover, wipe them clean with a soft 
dry cloth.
•  The operation with dirty gloves may cause a switch 
response error.
NOTE
The language on the display is set to “English” as default 
at the factory shipment. Select the adequate language 
referring to the Subsection 5.2.2. The menu pass of the 
display on this manual is selected to “English”.
5.2  Display and Basic 
Conguration
The display unit of AXW Integral Flowmeter and AXW4A/
AXG1A Remote transmitter has various functions below.
5.2.1 Display
microSD card
Status icon and time
Data display
(Abbreviation: 
see Table 5.2.1)
Switch On/Off status
(On: inverted)
IR switches
22.000mA
0.0%
0.00000
m/s
00 : 00
VEL
FLP
A01
SET SFT INC
(1)  Basic operation of IR switches
The operation from display panel is done by using 
the three IR switches; [SET], [SHIFT] and [▼]. The 
combination of the two switches provides a different 
function, and the function is indicated on the display.
IR switch  
(Note 1)
Indicate 
of switch 
(Note 2)
Function
[SET►] SET
▪ Apply parameter (Note 3)
▪ Enter data (Note 3)
▪ Move to next menu
[SHIFT] SFT
▪  Move cursor right 
(Numeric type parameter)
[▼] INC
▪  Move cursor down 
(Select type parameter)
▪  Increment value 
(Numeric type parameter)
▪  Change position of decimal point 
(Numeric type parameter)
[SHIFT] + [▼] 
(=[▲])
DEC
▪  Move cursor up 
(Select type parameter)
▪  Decrement value 
(Numeric type parameter)
SHIFT + SET► 
(=[ESC◄])
ESC
▪ Cancel
▪ Back to previous menu
Note 1:  [A] + [B] (=[C]): The function is changed to switch [C] 
when switch [B] is pushed while pushing switch [A].
Note 2:  [SET], [SFT], [INC], [DEC] and [ESC] indicate the 
assigned function in accordance with display mode at 
that time.
Note 3:  “Apply” and “Enter” are executed by pushing a switch 
twice. If a switch after rst pushing a switch.