7.3 List of Commands
Item Command Description Normal CA150
compatible
Data retrieval OD Queries the measurement value Ye s Yes
OE Queries the error information Yes Yes
OM Queries the memory data Yes Yes
OS Queries the settings (conditions) Yes Yes
Measurement
function
MR Sets or queries the measurement range Yes Yes
VO Starts, stops or queries the 24 V loop power supply Yes Yes
Source function SD Sets or queries the source value Yes Yes
SO Starts or stops or queries the source Yes Yes
SR Sets or queries the source range Yes Yes
Source/
measurement
settings
AG Sets or queries averaging Yes No
BU Sets or queries the burnout on/off state Yes No
MF Sets or queries the measurement function Yes Yes
MH Sets or queries the value corresponding to 100% of
FUNCTION1
Yes No
ML Sets or queries the value corresponding to 0% of
FUNCTION1
Yes No
MSC Sets or queries the FUNCTION1 display Yes No
PC Sets or queries the contact I/O Yes No
PU Sets or queries the PULSE (source) display Yes Yes
SF Sets or queries the source function Yes Yes
SH Sets or queries the value corresponding to 100% of
FUNCTION2
Yes No
SL Sets or queries the value corresponding to 0% of
FUNCTION2
Yes No
SP Sets or queries the pulse count operation state Ye s No
SSC Sets or queries the FUNCTION2 display Yes No
TC Sets or queries the pulse count time Yes Yes
TE Sets or queries the TC, RTD (source) display Yes Yes
WC Sets or queries the wiring system of resistance
measurement
Yes No
Device settings AP Sets or queries the auto power-off feature Yes No
DT Sets or queries the date and time Yes Yes
IO Sets or queries the on/off state of the 250 Ω internal
resistor
Yes Yes
LG Sets or queries the language Yes No
Other *IDN? Queries the CA500 or CA550 ID string Yes No
BGD Queries the calibration date of the CA500 or CA550 Yes No
BL Sets or queries the on/off state of the screen light Yes Yes
BSN Queries the serial number of the CA500 or CA550 Ye s No
DW Decreases the mth digit of the source value by 1. Yes Yes
ESC C/RC Initializes the settings Yes Yes
ESC S Outputs the instrument’s status byte. Yes Yes
H Sets or queries the output header of the OD and OM
commands
Yes Yes
HD Holds or queries the measurement value display Yes Yes
IM Sets or queries the status byte detection and mask Yes Yes
OR Queries the external RJ sensor connection Yes Yes
TS Executes manual saving Yes No
TT Sets or queries the international temperature standard Yes Yes
UP Increases the mth digit of the source value by 1. Yes Yes
YC Initializes the settings of FUNCTION1 and FUNCTION2 Yes No

Table of Contents

Other manuals for YOKOGAWA CA550

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA CA550 and is the answer not in the manual?

YOKOGAWA CA550 Specifications

General IconGeneral
BrandYOKOGAWA
ModelCA550
CategoryTest Equipment
LanguageEnglish

Related product manuals