EasyManua.ls Logo

YOKOGAWA CA550 - Page 111

YOKOGAWA CA550
125 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7-8
IM CA500-01EN
IO Sets or queries the on/off state of the 250 Ω internal resistor
Command = IOm<CRLF> -> Answer = IOm<CRLF>
Command = IO?<CRLF> -> Return = IOm<CRLF>
Parameters
m = 0: OFF
1: ON
LG Sets or queries the language
Command = LGm<CRLF> → Answer = LGm<CRLF>
Command = LG?<CRLF> → Return = LGm<CRLF>
Parameters
m = 0: English
m = 1: Japanese
m = 2: Chinese (Simplified)
m = 3: Korean
m = 4: Russian
MF Sets or queries the measurement function
Command = MFn<CRLF> -> Answer = MFn<CRLF>
Command = MF?<CRLF> -> Return = MFn<CRLF>
Command = MFm,n<CRLF> -> Answer = MFm,n<CRLF>
Command = MFm?<CRLF> -> Return = MFm,n<CRLF>
Parameters
m = 0:FUNCTION1
1: FUNCTION2 (for TC)
n = Function
0: DCV
1: DCA
2: Ω
3: TC
4: RTD
5: Freq
7: OFF
When the parameter is omitted, m=0 is assumed.
MH Sets or queries the value corresponding to 100% of FUNCTION1
Command = MHm<CRLF> -> Answer = MHm<CRLF>
Command = MH?<CRLF> -> Return = MHm<CRLF>
Parameters
m = 100% value
The setting range and resolution are the same as the display range of the selected range.
ML Sets or queries the value corresponding to 0% of FUNCTION1
Command = MLm<CRLF> -> Answer = MLm<CRLF>
Command = ML?<CRLF> -> Return = MLm<CRLF>
Parameters
m = 0% value
The setting range and resolution are the same as the display range of the selected range.
7.4 Commands