<8.ParameterSummary>
45
IM01C25A01-01E
No. Parametername Item
R/W
*1
Content Defaultvalue
Applicablemodel
Upload
data
F P L
J ADJUST Adjustingdata
J09 ADJUNIT Pressureadjustingunit
select
W %orPRESUNIT PRESUNIT
—
J10 ADJPRES Adjustmentreference
pressure
R UnitspeciedinJ09
—
J11 PZEROADJ Automaticzero
adjustment
W -32000to32000,unit
speciedinJ09
0.00000kPa
—
J12 PSPANADJ Automaticspan
adjustment
W -32000to32000,unit
speciedinJ09
100.000kPa
—
J15 PZERODEV Manualzeroadjustment W -32000to32000,unit
speciedinJ09
100.000kPa
—
J16 PSPANDEV Manualspan
adjustment
W -32000to32000,unit
speciedinJ09
100.000kPa
—
J20 ADJSP Adjustmentreference
pressureofSP
R UnitspeciedinJ09
—
—
J21 SPZEROADJ AutomaticSPzero
adjustment
W -32000to32000,unit
speciedinJ09
0.00000MPa
—
—
J22 SPSPANADJ AutomaticSPspan
adjustment*
2
W -32000to32000,unit
speciedinJ09
—
—
J25 SPZERODEV ManualSPzero
adjustment
W -32000to32000,unit
speciedinJ09
0.00000MPa
—
—
J26 SPSPANDEV ManualSPspan
adjustment
W -32000to32000,unit
speciedinJ09
0.00000MPa
—
—
J40 OUTPUT4mA 4mAadjustment W -10.000to10.000% 0.000%
—
J41 OUTPUT20mA 20mAadjustment W -10.000to10.000% 0.000%
—
J45 AMPTEMP Ampliertemperature R UnitspeciedD40 0degC
—
J50 ADJWHO Adjustmentinformation W 8alphanumericcharacters
—
J51 ADJDATE Adjustmentinformation W 16alphanumericcharacters
—
J52 ADJLOC Adjustmentinformation W 8alphanumericcharacters
—
J53 ADJDESC Adjustmentinformation W 16alphanumericcharacters
—
J55 EXTZEROADJ Externalzeroing
permission
W INHIBITorENABLE ENABLE
—
J56 CLEARADJ Clearadjustment W ---,PRES,SP,4-20mA,or
ALL
---
—
J60 SELFCHECK Self-diagnostics R SeeA60
—
K TEST Testparameters
K10 OUTPUTX% Testoutput%setting W Withinarangebetween
D20andD21
0.00%
—
K40 DOTEST Testcontactoutput W OFForON OFF
—
K45 TESTTIME “OUTPUTX%”and
“DOTEST”duration
timeselection
W 10min,30min,60min,
3hour,6hour,12hour
10min
—
K50 TESTKEY1 Specialmaintenance
parameter
W
—
K51 TESTKEY2 Specialmaintenance
parameter
W
—
K52 TESTKEY3 Specialmaintenance
parameter
W
—
K53 TESTKEY4 Specialmaintenance
parameter
W
—
K60 SELFCHECK Self-diagnostics R SeeA60
—
M DEVICEINFO Deviceinformation
M10 SERIALNO. Serialnumber R
—
M11 MFTR.DATE Manufactureddate R
—
M12 EXTRANO. Customizationnumber R
—
M15 SOFTREV Softwarerevision R
—
M16 BRAINREV BRAINprotocolrevision R
—
M17 MEMO1 Memo W 16alphanumericcharacters
—
M18 MEMO2 Memo W 16alphanumericcharacters
—
M19 MEMO3 Memo W 16alphanumericcharacters
—
*1: R/W:R=Readonly,W=Read&Write
*2: ThedefaultvalueshowsMWP(Maximumworkingpressure)ofthecapsule.
SincetheworkingpressurelimitvariesaccordingtotheModel,refertotheGeneralSpecicationssectionineachuser’smanual.
ParameterSummary