<8.ParameterSummary>
45
IM01C25A01-01E
No. Parametername Item
R/W
*1
Content Defaultvalue
Applicablemodel
Upload
data
F P L
J ADJUST Adjustingdata
J09 ADJUNIT Pressureadjustingunit
select
W %orPRESUNIT PRESUNIT
J10 ADJPRES Adjustmentreference
pressure
R UnitspeciedinJ09
J11 PZEROADJ Automaticzero
adjustment
W -32000to32000,unit
speciedinJ09
0.00000kPa
J12 PSPANADJ Automaticspan
adjustment
W -32000to32000,unit
speciedinJ09
100.000kPa
J15 PZERODEV Manualzeroadjustment W -32000to32000,unit
speciedinJ09
100.000kPa
J16 PSPANDEV Manualspan
adjustment
W -32000to32000,unit
speciedinJ09
100.000kPa
J20 ADJSP Adjustmentreference
pressureofSP
R UnitspeciedinJ09
J21 SPZEROADJ AutomaticSPzero
adjustment
W -32000to32000,unit
speciedinJ09
0.00000MPa
J22 SPSPANADJ AutomaticSPspan
adjustment*
2
W -32000to32000,unit
speciedinJ09
J25 SPZERODEV ManualSPzero
adjustment
W -32000to32000,unit
speciedinJ09
0.00000MPa
J26 SPSPANDEV ManualSPspan
adjustment
W -32000to32000,unit
speciedinJ09
0.00000MPa
J40 OUTPUT4mA 4mAadjustment W -10.000to10.000% 0.000%
J41 OUTPUT20mA 20mAadjustment W -10.000to10.000% 0.000%
J45 AMPTEMP Ampliertemperature R UnitspeciedD40 0degC
J50 ADJWHO Adjustmentinformation W 8alphanumericcharacters
J51 ADJDATE Adjustmentinformation W 16alphanumericcharacters
J52 ADJLOC Adjustmentinformation W 8alphanumericcharacters
J53 ADJDESC Adjustmentinformation W 16alphanumericcharacters
J55 EXTZEROADJ Externalzeroing
permission
W INHIBITorENABLE ENABLE
J56 CLEARADJ Clearadjustment W ---,PRES,SP,4-20mA,or
ALL
---
J60 SELFCHECK Self-diagnostics R SeeA60
K TEST Testparameters
K10 OUTPUTX% Testoutput%setting W Withinarangebetween
D20andD21
0.00%
K40 DOTEST Testcontactoutput W OFForON OFF
K45 TESTTIME “OUTPUTX%”and
“DOTEST”duration
timeselection
W 10min,30min,60min,
3hour,6hour,12hour
10min
K50 TESTKEY1 Specialmaintenance
parameter
W
K51 TESTKEY2 Specialmaintenance
parameter
W
K52 TESTKEY3 Specialmaintenance
parameter
W
K53 TESTKEY4 Specialmaintenance
parameter
W
K60 SELFCHECK Self-diagnostics R SeeA60
M DEVICEINFO Deviceinformation
M10 SERIALNO. Serialnumber R
M11 MFTR.DATE Manufactureddate R
M12 EXTRANO. Customizationnumber R
M15 SOFTREV Softwarerevision R
M16 BRAINREV BRAINprotocolrevision R
M17 MEMO1 Memo W 16alphanumericcharacters
M18 MEMO2 Memo W 16alphanumericcharacters
M19 MEMO3 Memo W 16alphanumericcharacters
*1: R/W:R=Readonly,W=Read&Write
*2: ThedefaultvalueshowsMWP(Maximumworkingpressure)ofthecapsule.
SincetheworkingpressurelimitvariesaccordingtotheModel,refertotheGeneralSpecicationssectionineachuser’smanual.
ParameterSummary
8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA EJX Series and is the answer not in the manual?

YOKOGAWA EJX Series Specifications

General IconGeneral
BrandYOKOGAWA
ModelEJX Series
CategoryTransmitter
LanguageEnglish

Related product manuals